Minimum requirements for monitors under the quality assurance standard (QS-RL)

Region of the body/ Examination Max. Display-
luminance in cd/m2
Min. luminance in cd/m2
*1
Max. luminance-
relationship
Matrix des BWGs
Pixel(w) x Pixel(h)
luminance-
curve
Homogeneity within BWG (between several BWG) Pixel error (each 1024 × 1024 pixel) Recommended JVC Display
Thorax ≥ 250 ≥ 1,1 × Lamb ≥ 250 ≥1600 × ≥1200 GSDF ± 10 % < 25 % (< 20 %) A ≤ 1
B ≤ 5
C ≤ 5
D ≤ 1
CL-S200
CL-S300
CCL214
CCL258i2
CCL358i2
CCL650i2
MS25i2
MS35i2
Extremities, skull Overview ≥ 150 ≥ 1,1 × Lamb ≥ 100 ≥1024 × ≥1024 GSDF ± 15 % < 25 % (< 20 %) A ≤ 1
B ≤ 5
C ≤ 5
D ≤ 1
CCL242
CL-S200
CL-S300
CCL214
CCL258i2
CCL358i2
CCL650i2
MS25i2
MS35i2
CT ≥ 150 ≥ 1,1 × Lamb ≥ 100 ≥1024 × ≥1024 GSDF ± 15 % < 25 % (< 20 %) A ≤ 1
B ≤ 5
C ≤ 5
D ≤ 1
CCL242
CL-S200
CL-S300
CCL214
CCL258i2
CCL358i2
CCL650i2
MS25i2
MS35i2
Für RK 5:
Dental digital volume tomography, intraoral X-ray diagnostics with dental tube devices, panoramic radiographs, cephalometric radiographs of the skull, dental skull overview images, hand surveys for determination of skeletal growth
≥ 200 ≥ 1,1 × Lamb ≥ 100 ≥1024 × ≥768 Entfällt
*2
< 30 % (< 30 %) A ≤ 1
B ≤ 5
C ≤ 5
D ≤ 1
CCL214
CCL242
Für RK 6:
Intraoral X-ray diagnostics with dental tube devices, panoramic radiographs, cephalometric radiographs of the skull, dental skull overview images, hand surveys for skeletal growth determination
≥ 300 ≥ 1,1 × Lamb ≥ 100 ≥1024 × ≥768 Entfällt
*2
< 30 % (< 30 %) A ≤ 1
B ≤ 5
C ≤ 5
D ≤ 1
CL-S200
CCL258i2
Mammography ≥ 250 ≥ 1,1 × Lamb ≥ 250 ≥2048 × ≥2048 GSDF ± 10 % < 25 % (< 10 %) A ≤ 1
B ≤ 5
C ≤ 5
D ≤ 1
MS55i2
CCL550i2
Mammography Lmax deviation ≥ 200 ≥ 1,1 × Lamb ≥ 100 ≥1024 × ≥1024 GSDF ± 15 % < 25 % (< 20 %) A ≤ 1
B ≤ 5
C ≤ 5
D ≤ 1
CCL242
CCL214
MS25i2

*1 In order to enable photopic vision, the value for the minimum luminance of 0.5 cd / m2 should not be significantly undercut.

* 2 Visual setting with TEST IMAGE OIQ a), b), c), e) and g) according to DIN 6868-157