DE

TOTOKU CCL 254i2

2 MP medical monitor for diagnostic imaging

TOTOKU CCL 254i2

The TOTOKU CCL254i2 high brighness monitor incorporates a range of sensors which collect, analyze and stabilize luminance, eliminating backlight fluctuation. Additionally, the backlight rise time achieves DICOM calibrated luminance levels in under a minute. This new generation of luminance technology is called λ-Sentinel II®.

The CCL252i2 features an extraordinary luminance of up to 950cd/m2, exceptional 170° viewing angles, is DICOM compliant, and includes an integrated hardware pivot function for a quick switch from landscape to portrait without the use of third party pivot software.

Product-Features

  • Medical Grade LCD-Monitor
  • Extra High Luminance: 950 cd/m² (Calibrated to 410 or 300 cd/m²)
  • Contrast Ratio: 900:1
  • Remote Calibration (PM Medivisor required)
  • Continuous DICOM Check (PM Medivisor required)
  • Uniformity Correction (Medivisor for i2 Series required)
  • Super fine Uniformity Correction
  • 5 Years Warranty incl. Backlight
  • Resolution: 1200 x 1600 / 2 Megapixels
  • Lambda-X3-Backlight
  • Luminance Stabilisation Lambda-Sentinel-II

Remote Calibration Capability

TOTOKU Uniformity Correction

The i2-models are equipped with an improved front sensor. Its accuracy opens up the opportunity of true remote calibration. The matching "PM-Medivisor" software is also available from TOTOKU.

It offers additional features, such as a continuous and automatic check of DICOM conformity.  This check is performed in the background unnoticed by the user and the results are reported to the user or the network administrator. All measured values are displayed in a clearly arranged graphic diagram.

 

 

The Uniformity Correction

TOTOKU Uniformity Correction

The i2-models feature a Uniformity Correction developed by TOTOKU. It involves subdividing the screen surface into multiple zones and performing extensive measurements on these zones. A combination of electronics and software is then used to adjust the brightness of the individual zones, much like in an equalizer, such that even minimal differences are eliminated in the process. This results in homogeneous luminance across the entire surface of the screen. The process can be repeated by the user at any time using the appropriate calibration software and a sensor. 

The Lambda Sentinel II

Lambda-Sentinel II Technology

The new TOTOKU i-Models have λ-Sentinel-II, an advanced automatic luminance control system that adjusts the luminance in perfection. The luminance sensor is places in front of the panel surface and stabilizes luminance by a permanent monitoring and sending feedback to the control circuit.

 

Calibration function to accurately adjust gamma, color temperature, and luminance

The calibration function creates the optimum conditions for a medical imaging display by adjusting luminance levels, color temperature, and grayscale characteristics to achieve DICOM GSDF compliant grayscale output.

more about the calibration function

LED Indicator

A glance at the LED indicator tells you the display's current operating status in three distinct colors. When in normal mode (luminance is stabilized), the LED is not lighted to minimize negatively impacting the diagnostic interpretation.

more about the LED Indicator

Worldwide valid safety standards for the medical field

TOTOKU monitors are completely certified according to the medical safety standards, which have more strictly demands than the safeties requirements for standard displays.

more about medical safety standards

Hardware Pivot Function

The pivot function allows for a 90-degree rotation of the LCD panel into portrait orientation.

more about the Hardware Pivot Function

 

Luminance Warranty for i2 models

Backlight warranty for i2 models

Backlight Warranty for i2 models
TOTOKU Models 5 Years or max.
operating hours
Calibrated
Luminance
cd/m²
Maximum
Luminance
cd/m²
Warranted
Luminance
in %
Reference
Luminance
in cd/m²
ME253i2 42.000h 410 1.800 35% 630
CCL252i2 42.000h 200 450 40% 180
CCL254i2 30.000h 410/300 950 40% 380
ME355i2 30.000h 410 1000 30% 300
CCL352i2 30.000h 250 600 35% 210
CCL354i2 30.000h 410/300 800 40% 320
ME551i2 30.000h 410 750 40% 300
MS21i2 42.000h 410 1.800 35% 630
MS51i2 30.000h 410 750 40% 300

Luminance warranty - further information


For all i2 models (from 01. November 2007) TOTOKU warrants the listed reference luminance in a warranty period of 5 years by a maximum of the listed operating hours.

  • 30.000 operating hours equal a daily usage of 16 hours on 365 days per year
  • 42.000 operating hours equal a daily usage of 23 hours on 365 days per year.


In spite of the high lifetime we recommend to switch off the display for longer breaks. So your can achieve an expanding lifetime for your monitors.

Graphic cards 2MP for color displays

i2-Product Files

All-In-One-Computer - Thin Client - Display

VARIO 322 medical All-In-One-Computer, Thin Client or Display

All-In-One-Computer - cost-efficient and hygienic

CLINIO 219C medical All-In-One-Computer, cost-efficient and hygienic

Now available with behind glass OSD

Imprint  |  § Terms & Conditions
 |  Telefon: +49 2156 4949-0
 |  Last Update: 16.09.2011 08:30